GB/T 1962.2-2001 Conical fittings with a 6%(Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings (English Version)
1 Scope
This standard specifies requirements for conical lock fittings with a 6% (Luer) taper for use with hypodermic syringes and needles and with certain other apparatus for medical use, e.g. transfusion equipment.
The requirements of this standard apply to fittings made of rigid and of semi-rigid materials and include test methods, but exclude provision for more flexible or elastomeric materials.
Note 1: It is not practicable to define the characteristics of rigid or semi-rigid materials with precision, but glass and metal may be considered as typical rigid materials. In contrast many plastics materials may be regarded as semi-rigid.
Note 2: The Luer lock fitting was designed for use at pressures of the order of 300 kPa or lower.
2 Normative References
The following standards contain provisions which, through reference in this text, constitute provisions of this standard. At the time of publication, the editions indicated were valid. All standards are subject to revision, and parties to agreements based on this standard are encouraged to investigate the possibility of applying the most recent editions of the standards indicated below.
GB/T 1031-1995 Surface Roughness Parameters and Their Values
GB/T 1962.1-2001 Conical Fittings with a 6% (Luer) Taper for Syringes, Needles and Certain Other Medical Equipment — Part 2: General Requirements (idt ISO 594-1:1986)
GB 15810-2001 Sterile Hypodermic Syringes for Single Use (eqv ISO 7886-1:1993)
GB/T 1962.2-2001 Conical fittings with a 6%(Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings (English Version)
Standard No.
GB/T 1962.2-2001
Status
valid
Language
English
File Format
PDF
Word Count
3000 words
Price(USD)
60.0
Implemented on
2002-2-1
Delivery
via email in 1 business day
Detail of GB/T 1962.2-2001
Standard No.
GB/T 1962.2-2001
English Name
Conical fittings with a 6%(Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
1 Scope
This standard specifies requirements for conical lock fittings with a 6% (Luer) taper for use with hypodermic syringes and needles and with certain other apparatus for medical use, e.g. transfusion equipment.
The requirements of this standard apply to fittings made of rigid and of semi-rigid materials and include test methods, but exclude provision for more flexible or elastomeric materials.
Note 1: It is not practicable to define the characteristics of rigid or semi-rigid materials with precision, but glass and metal may be considered as typical rigid materials. In contrast many plastics materials may be regarded as semi-rigid.
Note 2: The Luer lock fitting was designed for use at pressures of the order of 300 kPa or lower.
2 Normative References
The following standards contain provisions which, through reference in this text, constitute provisions of this standard. At the time of publication, the editions indicated were valid. All standards are subject to revision, and parties to agreements based on this standard are encouraged to investigate the possibility of applying the most recent editions of the standards indicated below.
GB/T 1031-1995 Surface Roughness Parameters and Their Values
GB/T 1962.1-2001 Conical Fittings with a 6% (Luer) Taper for Syringes, Needles and Certain Other Medical Equipment — Part 2: General Requirements (idt ISO 594-1:1986)
GB 15810-2001 Sterile Hypodermic Syringes for Single Use (eqv ISO 7886-1:1993)