GB/T 5526-2024 Animal and vegetable fats and oils - Determination of relative density
1 Scope
This document describes the digital density meter method (Method I), relative density meter (hydrometer) method (Method II), density bottle method (Method III), and density meter method (Method IV) for the determination of relative density of animal and vegetable fats and oils.
This document is applicable to the determination of the relative density of animal and vegetable oils and fats. Method I is applicable to the determination of the relative density of liquid and semi-solid animal and vegetable fats and oils; Methods II, III, and IV are applicable to the determination of the relative density of liquid animal and vegetable fats and oils.
2 Normative references
There are no normative references in this document.
3 Terms and definitions
No terms and definitions are listed in this document.
4 Digital density meter method (Method I)
4.1 Principle
Using the U-shaped tube oscillation method, lead the specimen into a U-shaped tube fixed at one end, and apply an excitation device to enable the U-shaped tube to oscillate at a specific characteristic frequency. The magnitude of the square of the oscillation frequency exhibits a linear relationship with the density or relative density of the specimen in the U-shaped tube.
4.2 Instruments
The digital density meter shall be equipped with automatic viscosity correction and temperature thermal equilibrium functions, and its accuracy shall comply with the following requirements:
—— temperature: (0±0.05)℃ ~ (100±0.05)℃;
—— relative density: (0±0.0001) ~ (3±0.0001).
If an autosampler is employed, the viscosity of the sample within the determination temperature range shall not exceed 35 Pa·s.
4.3 Steps
4.3.1 Calibrate the instruments in accordance with the requirements of the instrument manual prior to determination.
4.3.2 Determination of the relative density of liquid oil or fat: set the U-tube temperature to 20 ℃, uniformly inject the specimen with a syringe to fill a clean, dry U-tube, allow the temperature to equilibrate and stabilize at (20±0.05) ℃, confirm the absence of bubbles in the U-tube by means of the bubble detection function or through the observation window, and record the relative density value (d2020) once the instrument displays it stably.
Contents
Foreword i
1 Scope
2 Normative references
3 Terms and definitions
4 Digital density meter method (Method I)
5 Relative density meter (hydrometer) method (Method II)
6 Density bottle method (Method III)
7 Density meter method (Method IV)
Annex A (Informative) Distilled water density
GB/T 5526-2024 Animal and vegetable fats and oils - Determination of relative density
1 Scope
This document describes the digital density meter method (Method I), relative density meter (hydrometer) method (Method II), density bottle method (Method III), and density meter method (Method IV) for the determination of relative density of animal and vegetable fats and oils.
This document is applicable to the determination of the relative density of animal and vegetable oils and fats. Method I is applicable to the determination of the relative density of liquid and semi-solid animal and vegetable fats and oils; Methods II, III, and IV are applicable to the determination of the relative density of liquid animal and vegetable fats and oils.
2 Normative references
There are no normative references in this document.
3 Terms and definitions
No terms and definitions are listed in this document.
4 Digital density meter method (Method I)
4.1 Principle
Using the U-shaped tube oscillation method, lead the specimen into a U-shaped tube fixed at one end, and apply an excitation device to enable the U-shaped tube to oscillate at a specific characteristic frequency. The magnitude of the square of the oscillation frequency exhibits a linear relationship with the density or relative density of the specimen in the U-shaped tube.
4.2 Instruments
The digital density meter shall be equipped with automatic viscosity correction and temperature thermal equilibrium functions, and its accuracy shall comply with the following requirements:
—— temperature: (0±0.05)℃ ~ (100±0.05)℃;
—— relative density: (0±0.0001) ~ (3±0.0001).
If an autosampler is employed, the viscosity of the sample within the determination temperature range shall not exceed 35 Pa·s.
4.3 Steps
4.3.1 Calibrate the instruments in accordance with the requirements of the instrument manual prior to determination.
4.3.2 Determination of the relative density of liquid oil or fat: set the U-tube temperature to 20 ℃, uniformly inject the specimen with a syringe to fill a clean, dry U-tube, allow the temperature to equilibrate and stabilize at (20±0.05) ℃, confirm the absence of bubbles in the U-tube by means of the bubble detection function or through the observation window, and record the relative density value (d2020) once the instrument displays it stably.
Contents of GB/T 5526-2024
Contents
Foreword i
1 Scope
2 Normative references
3 Terms and definitions
4 Digital density meter method (Method I)
5 Relative density meter (hydrometer) method (Method II)
6 Density bottle method (Method III)
7 Density meter method (Method IV)
Annex A (Informative) Distilled water density